Maintenance And Cleaning; Routine Maintenance - Philips HeartStart FR3 Guía De Configuración, Funcionamiento Y Mantenimiento

Ocultar thumbs Ver también para HeartStart FR3:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

WARNING: Most cardiac arrests in children are not caused by heart problems. When responding
to cardiac arrest in an infant or child:
Alternatively, follow your local protocol.
3.
Leave the pads connected and the defibrillator turned on for continued monitoring and rhythm
analysis, as appropriate. Continue to follow any voice and screen prompts the FR3 provides.
NOTE: The FR3 automatically runs periodic self-tests. While a self-test is running, the green Ready
light is on steady, and the display is dark. If you press the On/Off button or open the FR3 carry case,
the FR3 aborts the self-test and turns on for use.

MAINTENANCE AND CLEANING

ROUTINE MAINTENANCE

In addition to the self-test performed each time it is turned on and each time a battery is installed, the
FR3 also automatically performs periodic self-tests. These include daily, weekly, and monthly self-tests
of respectively increasing detail to verify readiness for use.
NOTE: The FR3 self-tests are designed to check that the FR3 is ready for use. However, in the event
that the FR3 has been dropped or mishandled, Philips recommends that you run a user-initiated test.
If the FR3 has visible signs of damage, contact Philips at www.philips.com/AEDSupport for technical
support.
If the green Ready light is blinking and the FR3 is not chirping, the FR3 has passed its most recent self-
test and is therefore ready for use. The defibrillator requires no calibration or verification of energy
delivery. Routine maintenance is limited to periodically conducting the following steps:
 Check the green Ready light. If the green Ready light is not blinking or the FR3 is chirping, see
 Replace any expired supplies and accessories.
 Check the outside of the defibrillator. If you see signs of damage, contact Philips for technical
CAUTION: Electrical shock hazard. Dangerous high voltages and currents are present. Do not open
the defibrillator, remove its covers, or attempt repair. There are no user-serviceable components in
the FR3. Return the FR3 to Philips for repair.
10
• Provide infant/child CPR while a bystander calls EMS and brings the FR3.
• If no bystander is available, provide 1-2 minutes of CPR before calling emergency medical systems
and retrieving the FR3.
• If you witnessed the child's collapse, call EMS immediately and then get the FR3.
If prompted by the FR3, PRESS THE SHOCK BUTTON to deliver a shock.
Troubleshooting, below.
support.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

861388861389

Tabla de contenido