Guida e dichiarazione del costruttore – IMMUNITÀ ELETTROMAGNETICA – PER
GLI APPARECCHI ED I SISTEMI CHE NON SONO DI SOSTENTAMENTO DI FUNZIONI
Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY – FOR
EQUIPMENT AND SYSTEMS THAT ARE NOT LIFE-SUPPORTING
Il MIO-PERISTIM è previsto per funzionare nell'ambiente elettromagnetico sotto
specificato. Il cliente o l'utilizzatore del MIO-PERISTIM deve garantire che esso
venga usato in tale ambiente.
MIO-PERISTIM is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in such
an environment.
broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which MIO-
PERISTIM is used exceeds the applicable RF compliance level above, MIO-PERISTIM
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating MIO-
PERISTIM.
b)
L'intensità di campo nell'intervallo di frequenza da 150 kHz a 80 MHz dovrebbe essere
minore di 3 V/m / Over the frequency range 150kHz to 80MHz, field strengths should be
less than 3V/m.
IACER Srl
VITALI
200