MIO-PERISTIM PHYSIO does not allow the intensity adjustment or not
•
keep the adjusted value and reset:
−
It is suggested to replace the battery and start a new treatment.
If the problem persists contact the manufacturer
Battery charging
MIO-PERISTIM is supplied by internal rechargeable Ni-MH 800mAh battery
with new long-lasting technology.
When during the treatment many intensity increases is needed or the device
turns off, it indicates a low battery state. In this case, the display will show
low battery indicator
the therapy session, or not being able to complete it.
To proceed with the charging follow the steps below:
•
make sure that the device is switched off before charging with the
button;
make sure that the device is NOT being used by patient (disconnect
•
cables and electrodes);
connect the battery charger to the plug on the upper side of MIO-
•
PERISTIM and connect the battery charger into the power socket.
The display will show the battery blinking icon and the TIME-min icon (which
takes account of the charging time) on the display. After 4 hours the recharge
automatically finishes and the display shows the recharge total time.
At the end of battery charging, disconnect the charger from power supply and
store it in the carriage bag.
WARNING: at the end of the charge wait at least 30 minutes before
switching on the device; in order to allow the cooling of the battery
pack, overheated during charging and the closure of the integrated
safety system that prevents the device from turning on.
Do not immerge the battery in water or other liquids and do not expose it to
heat sources.
Do not dispose of dead or defective batteries with domestic waste; dispose of
in an authorized waste collection bin or in any case according to the
underlying norm (WEEE).
Only adults should be managing the battery. Keep out of children's reach.
Do not use the battery charger if:
•
the plug is damaged or parts of it are broken;
•
it has been exposed to rain or any other type of liquid;
•
the components have been damaged by a fall.
IACER Srl
. In this case it may not be possible to undertake
70
MNPG152-07