It is Recommended to Perform a Control Solution Test:
• Whenever you suspect the meter or test strips may not be functioning
properly
• If test results appear to be abnormally high or low or are not consistent
with clinical symptoms
• The test strip bottle has been left open or has been exposed to
temperatures below 34°F (1°C) or above 86°F (30°C), or humidity levels
outside the range of 20–80% relative humidity
• To check your technique
• When the ReliOn™ Confirm blood glucose meter has been dropped or
stored below 32°F (0°C) or above 122°F (50°C)
• When you open a new bottle of test strips
IMPORTANT:
Use only with ReliOn™ Confirm blood glucose meter and ReliOn™
Confirm/Micro Plus test strips.