OBJ_DOKU-192307-004.fm Seite 18 Mittwoch, 1. April 2020 4:42 16
1.4

Qualification of user

For the intended use, the product may only be operated by med-
ical professionals who have been trained in its use on the basis
of the notes on use.
1.5

Performance characteristics

Performance characteristics relating to the intended use are:
• Transporting HF current from the electrosurgical unit to the
instrument.
2

Maximum electrical capacity

The maximum electrical capacity of this product is:
• 1000 V
p
3

Safety instructions

WARNING!
Check insulation for damage before every application.
Erbe recommends carrying out an electrical continuity test be-
fore each application of these products.
Note: If the cable conductor is broken or the electrical continuity
of the cable is interrupted in any other manner it can cause arc-
ing in the patient return system or in the active circuit; this can
lead to the patient suffering burns or the outbreak of a fire.
If damaged, do not use these products!
Do not use in the presence of combustible or explosive materi-
als!
Never lay this product on the patient or in his/her direct vicinity.