Replacing The Cartridge - NSK Pana-Max2 M4 Manual De Operación

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 42
• I mmediately after use, the product should be cleaned, lubricated
and sterilized. If blood remains on the external or internal surfaces
it can become clotted and cause rust.
• D o not heat or cool the product too quickly. Rapid change in
temperature could cause damage to the product.
• I f the sterilizer chamber temperature may exceed 135°C during
the dry cycle then delete the dry cycle.
• A utoclave sterilization is recommended for the product. The
validity of other sterilization methods is not confirmed.
• D o not touch the product immediately after autoclaving as it will be
very hot and must remain in a sterile condition.
NOTICE
• N SK recommends Class B sterilizers as stated in EN13060.
8

8. Replacing the Cartridge

1) Insert a test bur.
2) Locate the correct wrench tool on the head cap then turn the
wrench counter clockwise to loosen the cap. Remove the cap.
3) Use the bur to gently lever the entire cartridge out from the head.
4) Clean the head interior with NSK PANA SPRAY Plus.
5) Wipe the NSK PANA SPRAY Plus oil over the head interior.
6) Insert the new cartridge into the head by aligning the pin on the
cartridge with the slot on the head. (Fig.15)
7) Firmly tighten the head cap with the correct head cap wrench.
*Refer to Spare Parts List to identify the correct cartridge.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Pana-max2 b2Pana-max2 qdPana-max2 ptlPana-max2 kv

Tabla de contenido