· Do not wash, soak, or wipe off the handpiece with/in the oxidation potential solution (strong acid, super acid solution) sterilized solution, strong
alkaline solution or functional water.
· The handpiece is kept in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. Moreover, air does not include a dust,
salt and sulfur.
· Do not touch the handpiece just after the autoclave. It may very hot.
· NSK do not recommend you to sterilize on plasma sterilization or EOG sterilization.
6. Replacing the Cartridge
1 ) Insert a dummy bur into the chuck.
2 ) Unscrew the Head Cap by turning counter-clockwise and remove.
3 ) Push up the dummy bur upwards to remove Cartridge.
4 ) Clean the interior of the Head casing.
5 ) Insert the new Cartridge.
6 ) Replace the Head Cap and firmly screw into place.
*SX-PU03: Order Code P1140200
14
Head
PUSH
Cartridge
Head Cap
Wrench
O-Ring
Fig. 12