Remote Sampling (With Pump Option); Pump Option - GMI PS500 Manual Del Usuario

Ocultar thumbs Ver también para PS500:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 79
USER HANDBOOK

2.7 REMOTE SAMPLING (with pump option)

Warning: When making VOC measurements, only use the
Viton sample line as other sample lines may
absorb VOC's resulting in incorrect readings.
Remote sensing is performed with the internal electric pump
(optional), or by the hand aspirator for non-reactive gases,
using the sample connector at the bottom of the instrument
and sample tubing supplied with your instrument.
On pumped models the pump is OFF after start-up.
Warning: For remote sampling, the internal pump is
recommended. If using the hand aspirator:
Reading error of + 20% is typical
Maximum sample line is 10 m
Sample time will be extended

2.7.1 Pump Option

Press and hold the
pump will run at normal speed for remote sampling.
When the pump Is running at normal speed, a pump symbol
rotates in the display as shown in Fig. 2.17.
2-14
button to start or stop the pump. The

Publicidad

Tabla de contenido
loading

Productos relacionados para GMI PS500

Tabla de contenido