Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain
a device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to
maintain a device may compromise the clinical condition or safety of users and/or their caregivers. Do not
perform service and maintenance while the user is in the product.
1st Day
EN
•
Read the User Guide carefully before use and
store it in a safe place for future reference
•
Store the enclosed tools on/at the product
Daily
•
Use a dry cloth to clean the product
•
Visual inspection. Repair or replace any
damaged or worn-out parts
•
Ensure that all fasteners are in place and fitted
correctly
•
Check all fittings and buckles for signs of worn
out parts
•
Be aware of any MAX mark indications before
adjustment
•
Check that all cables connecting the power
system are in good shape
Weekly
•
Check that all castors are moving freely and
that all wheel locks/brakes are easily activated
•
Wipe off any residue and dirt from the product
using a cloth with warm water and a mild
detergent/soap and dry before use
•
Disinfect the product. please check the
disinfection section in this User Guide for more
information
Monthly
•
All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
•
Oil the swing-away parts. We recommend you
use a professional lubricant system
Annually
•
Inspect the frame for cracks or signs for worn-
out parts and carry out annual service. Never
use a product which appears to be substandard
or faulty
Warewashing machine
This product can be washed at:
•
60°C in 10 minutes
•
85°C in 3 minutes
.. using a mild detergent in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product. Clean the
accessories separately.
Before washing
Remove the following items and clean them
separately:
•
Accessories
•
Power unit including; Actuator, battery and
handset
After washing:
•
Replace the power column going backwards
through the procedure give on page 14
•
Check that the handset functionalities works
correct
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to wipe
off any residue and dirt from the product, using a
cloth with warm water and a mild detergent/soap
without chlorine and let it dry before disinfection.
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months for heavy users)
and every time the product is reissued for use. The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 5 years
if all maintenance and servicing is carried out in
accordance with the manufacturer's instructions
and demonstrably recorded.
Service Check List:
Check and adjust the following functions as per their
intended use. Repair or replace any damaged or
worn-out parts. Please check the:
•
For products with gasspring and/or telescopic
tube, check that the locking mechanisms work.
Also check that it is possible to tilt the basis
plate horizontally and vertically.
•
For products with power, recharge the battery
and check that the handset works by tilting the
basis plate to vertical and horizontal position
•
Check that all castors are moving freely and
that all wheel locks/brakes are easily activated
•
The foot plate(s) is adjustable, both distance
to basisplate and the angle mechanism must
be checked
•
All cushions are in good shape without cracks
and or holes in the cover
•
Nuts and bolts on the product are tightened
•
The straps must be without cracks and not
flossed, pay special attention to the area where
the straps connect to the brackets.
•
The plastic pieces connecting the frame to the
basis plate is undamaged and without cracks.
Pay special attention to the areas where
fasteners are inserted
•
The rails in the basis plate must be in good
shape, it should be possible to slide a strap up
and down the rail.
•
All labels on the product are intact