Power unit
For an optimum lifetime the power unit must
be connected to the mains voltages as often as
possible (e.g. every night). Charge the batteries 12
hours before use.
The batteries must be charged at least every 3rd
month - otherwise there might be damages due to
self-discharge. It is recommended to test the battery
function at least once every year.
Recharging the battery:
•
Plug the charging lead into the handset
•
Plug the charger into the electricity supply and
switch ON. A yellow light on the charger will
light up
•
Charge the batteries until the lamp on the
charger turns green
Caregiver
•
Do not leave children and youngsters alone
with the handset
•
Do not let children and youngsters play with the
handset
•
Do not leave the handset within reach of
children and youngsters.
Product
•
The power unit does not have a safety stop
•
The system must not be sprayed directly with a
highpressure cleaner
•
Cleaning with a steam cleaner is not permitted
•
Do only use the power unit within specified
working limits
•
It is recommended to check the handset and
cable for damage and holes caused by violent
handling at least once a year
•
Do not submerge the handset in water
•
Do not attempt to repair any part of the power
unit.
Environmental conditions
•
Operating temperature: +5°C to +40°C
•
Storage temperature: -10°C to +50°C
•
Relative humidity: 20% to 90%
•
Atmospheric pressure: 800 to 1060 hPa
support.R82.org
Technical Data
Battery - BA18
•
Capacity 1.3 Ah - 24V DC
•
Charging current is max. 500 mA
•
IPX6
•
Duty cycle: Max. 10% or 2 min. use then 18
min. without use
The batteries have a maximum life of 4 years but this
can vary depending upon the conditions of use and
charging practices
Battery charger - CH01
•
Mains: 100 - 240VAC/50-60 Hz switch mode
•
Charging voltage: 27.6 VDC
•
Charging current: Max. 500 mA
Actuator - LA28
•
Input voltage: 24V
•
Noise level: ≤ 45 dB(A)
•
IPX1
Handset - HB 40
•
IPX1
Detachable parts
•
Handset - HB 40
•
Battery - BA 18
Applied part
•
Handset - HB 40
IP-Rating table
No protection
IPX0
IPX1
Protection against vertically falling
drops of water e.g. condensation
IPX2
Protection against direct sprays of
water up to 15° from the vertical
IPX3
Protected against direct sprays of
water up to 60° from the vertical
Protection against water sprayed
IPX4
from all directions - limited ingress
permitted
Protected against low pressure jets
IPX5
of water from all directions - limited
ingress
IPX6
Protected against temporary flooding
of water, e.g. for use on ship decks -
limited ingress permitted
IPX7
Protected against the effect of im-
mersion between 15 cm and 1 m
IPX8
Protects against long periods of im-
mersion under pressure
27