Cleaning And Disinfecting - Permobil ROHO AGILITY Minimum/Mid Contour Back Support Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Cleaning and Disinfecting

Warnings:
- Cleaning and disinfecting are separate processes. Cleaning must precede
disinfection. Before use by a different individual: clean, disinfect, and check
product for proper functioning.
- Clean the product regularly and when it becomes soiled.
- If instructed to use bleach, follow the safety guidelines on the bleach
container.
Recommendation: When washing by hand, use room-temperature water.
Back Shell
To Clean: Hand wash with a sponge, using soap and water. Rinse with clean water. Wipe dry with a clean cloth and then allow to air dry
completely.
To Disinfect: Wipe gently with a cloth dampened with a household disinfectant. Rinse with clean water. Wipe dry with a clean cloth and
then allow to air dry completely.
Cover
Notes: Before cleaning and disinfecting, remove the foam and air components from the cover. Secure the hook and loop fastener strip on the handle at the wide
end of the cover. Turn the cover inside out. Zip the cover closed to avoid snags.
To Clean: Machine wash cold, with mild detergent, gentle cycle, or wipe clean with neutral detergent and water. Rinse
thoroughly. Tumble dry low or air dry.
To Disinfect: Hand wash, using quaternary ammonium disinfectant (follow product label instructions). Rinse
thoroughly with clean water. Or machine wash in hot (60°C) water. Tumble dry low or air dry. DO NOT use
bleach.
Foam pad, foam insert, and lumbar support pad
Note: Remove the foam components from the cover prior to cleaning.
To Clean: Wipe lightly with a damp cloth. DO NOT use soap. DO NOT submerge in water. Wipe off with a clean cloth. Completely air dry
before replacing in cover.
Warning: Disinfectants are NOT effective on porous surfaces such as foam. Replace soiled or contaminated foam components with authorized ROHO foam
components only. Replacement foam components may be purchased separately. Soiled or contaminated foam components MUST NOT be used by different
individuals.
Air Insert
Note: Remove the air insert from the cover and foam pad prior to cleaning and disinfecting.
Cautions: Make sure the inflation valve is closed. DO NOT allow water or cleaning solution to enter the air insert. DO NOT machine wash or
dry. DO NOT expose to direct sunlight. DO NOT autoclave. Rinse thoroughly. Detergent residue may cause air cells to stick together.
To Clean: Close valve. Place in a large sink. Hand wash, using mild liquid hand soap, hand dishwashing liquid, laundry detergent, or a
multipurpose detergent (follow product label instructions.) Gently scrub all cushion surfaces, using a soft, plastic bristle brush, a sponge, or a
cloth. Rinse with clean water. Air dry.
To Disinfect: Follow the cleaning instructions, using 1 part household liquid bleach per 9 parts water. Keep the product wet with the
bleach solution for 10 minutes. Rinse thoroughly with clean water. Air dry.
Note: Deflated air cells may be easier to thoroughly clean. To deflate the air cells: with the inflation valve open, roll up and/or compress the cushion, and then
completely close the inflation valve.
Care Symbols
Hand wash.
Dry flat.
Wipe with a damp cloth.
ROHO AGILITY Minimum/Mid Contour Back Support Operation Manual
30C
Machine wash cold, gentle.
Tumble dry low.
Dry flat in the shade.
Cautions:
- DO NOT use abrasives such as steel wool or scouring pads on components
of an AGILITY.
- DO NOT use cleaning products containing petroleum or organic solvents [i.e.
acetone, toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluids,
adhesive removers] on components of an AGILITY.
Machine wash hot, normal, at the temperature
60C
shown.
Do not bleach.
Bleach (1 part bleach : 9 parts water).
1: 9
9
30C
or
60C
or
or
+
1:9
Do not machine wash.
Do not tumble dry.
Line dry.

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido