3. Adjusting Radial Arm Swing Tension
To adjust radial arm swing tension:
1. Remove the radial arm cover. Using a 3/32" hex key, remove the two screws securing the top cover.
2. Insert a 5/32" hex key in the point shown in 3 above.
3. Turn the tensioning screw clockwise to add tension or counterclockwise to remove tension.
4. Replace the arm cover.
4. Adjusting Monitor Tilt Tension
If the monitor droops, adjust the tilt tension:
1. Tilt the monitor upward and insert a 7/32" hex key in the adjustment point shown.
2. Turn the tensioning screw clockwise or counterclockwise until the monitor no longer droops when positioned.
5. Adjusting Upper Arm Pivot Tension
To adjust the upper arm pivot tension:
1. Insert a 3/32" hex key at the point shown in 7 above.
2. Turn the tensioning screw clockwise to add tension or counterclockwise to remove tension.
P44082 Rev. D
EN 11