Press the ESC button.
Note: Before pressing ESC you can press the arrow buttons
to view test quality
messages,
predicted and measured
results, but this should not be necessary.
T e s t
# 1
c o m p l e t e
R e a d y
f o r
n e x t
a n d
↓
f o r
o r
E S C
t o
e n d
Performing an FVC Test
Select FVC in the Test menu. Give the device to the subject.
P E R F O R M
F V C
E x p i r a t o r y
Give the following instructions to the subject to ensure that
testing is performed properly, first demonstrating youself with
a mouthpiece (adding "like this"):
•
"I am now going to test how fast you can blow out"
•
"Inhale as deeply as possible, and hold your breath"
•
"Insert the mouthpiece into your mouth, biting it lightly"
•
"Close your lips round the mouthpiece".
•
Exhale as fast as possible, keep going, go on..."
•
Keep the subject exhaling for at least 6 seconds. An
animated performance by the operator is essential to
get the subject to perform a good test.
Testing will be accompanied by audible feedback if
Sound is selected in the Setup menu. If inspiratory
parameters are required, continue with:
"Now, fully inhale as quickly as possible"
The unit is ready to accept another test when two beeps
occur (if Sound is enabled). After resting time, re-test the
subject until you are satisfied that a good FVC has been
performed or until the unit beeps five times, which indicates
that the ATS criteria for test quality have been fulfilled or that
eight tests have been completed in this session.
Take back the spirometer and congratulate:
•
"Well done".
S u c c e s s f u l
s e s s i o n
T e s t i n g
P r e s s
↵
f o r
a n d
↓
f o r
Note: if, for some reason, it is required that testing continues,
simply press ESC to proceed.
The patient's best test can be recorded on a Serial
Spirometry Record sheet along with a plot of the FEV1 value.
Pads of 50 can be ordered from Vitalograph (Cat. No. 63054).
After performing the test series, return the device back to the
cradle.
Normally no examination of the Vitalogaph 2120 display is
required, but various information is displayed for the operator
who is having difficulties:
S l o w
s t a r t
o f
A b r u p t
e n d
o f
D u r a t i o n :
4 . 6
W A I T
P L E A S E . . .
Note: Quality messages like this will be shown on screen for
a few seconds. relating to the test just performed.
8
All manuals and user guides at all-guides.com
b l o w
r e s u l t s
t e s t s
T E S T
o n l y
o v e r
m e n u
o r
r e s u l t s
t e s t
t e s t
T e s t
# 1
c o m p l e t e
R e a d y
f o r
n e x t
a n d
↓
f o r
o r
E S C
t o
e n d
Press the
or
to move through a quality summary of the
test session to date and the pre-configured indices, predicted
and measured results.
L A S T
F V C
5 . 3 5
F E V 1
5 . 1 2
P E F
5 2 4
P
P
-B
ERFORMING
OST
RONCHODILATOR
Bronchodilation or provocation testing may be performed in
Post mode. Post tests can be performed on a subject who
has been previously tested. Select Post in the main menu.
Post testing can be performed on the current subject or can
be performed on a previously stored subject.
S e l e c t
P o s t
>
C U R R E N T
R E C A L L
P A T I E N T
Testing can continue as described previously.
V
T
R
IEWING
EST
ESULTS
The results for indices are displayed automatically when you
perform a test. You can also view the current test results or
previously stored results for all the indices selected in the
Setup menu. Subject demographics can also be recalled for
viewing. This achieved by selecting View in the main menu.
S e l e c t
V i e w
>
C U R R E N T
T E S T
R E C A L L
T E S T
P A T I E N T
D A T A
If selecting a test to view, then you will see a screen like this.
S e l e c t
T e s t
1 3 : 1 2
1 2 / 1 2 / 9 5
1 3 : 1 2
1 2 / 1 2 / 9 5
1 3 : 2 0
1 2 / 1 2 / 9 5
Press the
button to move the cursor to the test you wish
select.
Note: a test marked with 'p' signifies that it is a post test.
Subject demographics can be recalled for viewing by
selecting Patient Data and the appropriate subject .
P a t i e n t
D e m o g r a p h i c s
0 2 0 1 5 _ 2 1 1 2 9 5 P 3
S e x
M
H e i g h t
A g e
2 5
R a c e
P
T
R
RINTING
EST
ESULTS
The Vitalograph 2120 is capable of printing informative, well
presented reports. Ensure that the correct printer is selected
in the Setup menu as previously described.
There are three choices available for printing test results:
•
printing the results of the current test
•
printing the results of all stored tests for all subjects
•
printing the results of a particular test stored for a
specified subject
Connecting to a Printer
For printing, the Vitalograph 2120 Printer Cable (2120505) is
required to interface between the cradle and the parallel port
of the printer. Use the following procedure.
b l o w
r e s u l t s
t e s t s
B E S T
5 . 3 6
5 . 1 2
5 2 4
T
ESTS
T y p e
O p t i o n
V
<
F p
F
1 4 0
C a u c