Installation - Agilent Technologies Turbo-V 250 Manual De Instrucciones

Tabla de contenido

Publicidad

Installation

WARNING!
Connection to the mains must be made in accordance with the local law.
Always connect the ground wire and use a properly grounded power socket
to avoid electrical shock. High voltage developed in the controller can cause
severe injury or death. Before servicing the unit, disconnect the input power
cable.
The Turbo-V controller can be used as a bench unit or a rack module, but it must
NOTE
be positioned so that free air can flow through the holes. Do not install or use
the controller in an environment exposed to atmospheric agents (rain, snow,
ice), dust, aggressive gases, or in explosive environments or those with a high
fire risk.
During operation, the following environmental conditions must be
respected:
To connect the controller to the pump use the specific cable supplied
with the controller.
The controller model 969-9504 is not equipped with the pump cable. It can be
NOTE
requested as an accessory; detailed information is supplied in the paragraph
"Accessories and Spare Parts" of the appendix "Technical Information". It must
be connected between the controller connector J16 and the pump connector.
The two connectors at the two cable extremity cannot be connected in a wrong
way.
Turbo-V 250 Controller User Manual / 87-900-856-01 (E)
temperature: from 0 °C to +40 °C
relative humidity: 0 – 95 % (without condensation).
Instructions for Use

Installation

197/240
16