Agilent Technologies Turbo-V 250 Manual De Instrucciones página 226

Tabla de contenido

Publicidad

17
Technical Information
Use
D
where: XXX = YES or NO.
If YES is selected, R2 energizes from the start of the turbopump and
deenergizes after the run-up time when the pump rotational speed
becomes lower than the programmed threshold value.
If NO is selected, R2 energizes from the start of the yutbopump and,
after the rotational speed of the pump exceeds for the first time the
speed threshold value, R2 deenergizes once the pump speed becomes
lower than the programmed speed threshold value (see the preceding
cycle diagram).
This function is factory-set to YES.
D
Figure 21
226/240
When the last digit is entered, the display shows:
E
L
A
Y
T
H
R
E
Press the CYCLE NUMBER pushbutton and select YES or NO.
E
L
A
Y
T
H
R