Procedure
Open the "SETUP" / "BALANCE" menu.
t
If the calibration function needs to be set with subse-
t
quent automatic adjustment: Select the "ADJUST"
calibration value for the "CAL.SEQ." parameter.
If the calibration function needs to be set without
t
subsequent automatic adjustment: Select the
"CAL-ADJUST" calibration value for the "CAL.SEQ."
parameter.
Setting the Weight Value for the External Weight
A preset weight value or a user-defined weight value can
be set for the external weight.
7.3
Parameter List
7.3.1
"SETUP" / "BALANCE" Menu
Parameter
Setting values
AMBIENT
V.STABLE
STABLE*
UNSTABL.
V.UNSTBL.
* Factory setting
Procedure
If the preset weight value needs to be used: In the
t
"SETUP" / "BALANCE" menu, for the "CAL./ADJ."
parameter, select the "EXT.CAL." setting value.
If a user-defined weight value needs to be set:
t
t
t
t
Explanation
Sets the ambient conditions to "very stable": Activates a fast change in the
weight values in the event of a load change with a high output rate.
Recommended for the following work environment:
− Very stable table near the wall
− Closed and calm room
Sets the ambient conditions to "stable". Recommended for the following
work environment:
− Stable table
− Slight movement in the room
− Slight draft
Sets the ambient conditions to "unstable": Activates the delayed change in
weight values with a reduced output rate. Recommended for the following
work environment:
− Simple office desk
− Room with moving machinery or personnel
− Slight air movement
Sets the ambient conditions to "very unstable": Activates a significantly
delayed change in the weight values and long wait for stability with a further
reduction in the output rate.
Recommended for the following work environment:
− Noticeable and slow floor vibrations
− Noticeable building vibrations
− Weighed goods moved
− Very strong air movements
System Settings
In the "INPUT" menu, select the "CAL.WT." set-
ting value.
Enter the desired weight value and press the
[Confirm] button.
To use the user-defined weight value for the next
calibration: In the "SETUP" / "BALANCE" menu,
for the "CAL./ADJ." parameter, select the "E.CAL.
USR" setting value.
Entris
BCE Operating Instructions
®
21