Calibration Mode

GB
The result is shown in the display, alternating with CAL.
RESULT
CAL
If the reading corresponds with the value of the calibration standard
(within the specified tolerance), exit calibration mode by pressing the
[ON/OFF] key.
Changing the displayed value:
Pressing the [MODE] key once increases the displayed value by 1 digit.
Mode
Pressing the [ZERO/TEST] key once decreases the displayed value by
Zero
1 digit.
Test
Press the corresponding key until the reading equals the value of the
CAL
calibration standard.
RESULT + x
By pressing the [ON/OFF] key, the new correction factor is calculated
On
and stored in the user calibration software.
Off
:
:
Confirmation of calibration (3 seconds).
CAUTION:
Separate calibration of the measuring range for Ur2 is not possible. The
unit uses the calibration for the Ur1 measuring range.