Procedures: Mv/Volts Input; Procedures: Mv/Volts Output; Procedures: Hz Input/Output; Procedures: Cj Input - Baker Hughes Druck DPI 880 Manual Del Usuario

Ocultar thumbs Ver también para Druck DPI 880:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 45

Procedures: mV/Volts input

1. Connect the instrument to the calibration equipment
(Figure 4).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a three-point
calibration (-FS, Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable mV or Volts input task (Table 2).
5. Then apply the input values that are applicable to the
calibration:
• mV: -120, -60, -30, 0 (short circuit)
Then mV: 0, 30, 60, 120
OR
• Volts (V): -30, -15, -5, 0 (short circuit)
Then volts (V): 0, 5, 15, 30
6. Make sure that the error is in the specified limits (Table 12
or Table 13).
Table 12: mV input error limits
Applied
Calibrator
mV
error
(mV)
±120
0.001 3
±60
0.000 8
±30
0.000 6
0 (Short circuit)
-
Table 13: Volts (V) input error limits
Applied
Calibrator
V
error
(V)
±30
0.000 58
±15
0.000 11
±5
0.000 06
0 (Short circuit)
-

Procedures: mV/Volts output

1. Connect the instrument to the calibration equipment
(Figure 4).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a two-point
calibration (Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable mV or Volts output task (Table 2).
5. Then set the output values that are applicable to the
calibration:
• mV: 0, 30, 60, 90, 120
OR
• Volts (V): 0, 3, 6, 9, 12
6. Make sure that the error is in the specified limits (Table 14
or Table 15).
Table 14: mV output error limits
Output
Calibrator
mV
error
(mV)
0
0.000 05
30
0.000 425
60
0.000 8
90
0.001 175
120
0.000 98
12
- [EN] English
Permitted
DPI 880 error
(mV)
0.03
0.02
0.02
0.01
Permitted
DPI 880 error
(V)
0.004
0.002
0.001
0.001
Permitted
DPI 880 error
(mV)
0.01
0.02
0.03
0.03
0.04
Table 15: Volts (V) output error limits
Output
Calibrator
V
error
(V)
0
0.000 000 05
3
0.000 017 5
6
0.000 03
9
0.000 05
12
0.000 134

Procedures: Hz input/output

1. Connect the instrument to the calibration equipment
(Figure 6).
2. Let the equipment get to a stable temperature (minimum:
5 minutes since the last power on).
3. Set up the equipment with these conditions:
Frequency meter:
Signal generator:
DPI 880:
4. Use the calibration menu (Table 9) to do the calibration.
The display shows the applicable instructions to complete
the calibration.
5. To make sure that the calibration is correct, set up the
equipment to do one of these calibration checks:
• Hz input calibration check (Figure 6):
Frequency meter:
Signal generator:
DPI 880:
• Hz output calibration check (Figure 6):
Frequency meter:
DPI 880:
6. Measure or supply the specified values (Table 16/17): Hz
then kHz. Make sure that the error is in the specified
limits.
Table 16: Hz error limits (Measure/Supply)
Measure/
Calibrator
Supply
error (Hz)
Hz
25
0.000 175
100
0.000 7
250
0.001 75
500
0.003 5
990
0.006 93
Table 17: kHz error limits (Measure/Supply)
Measure/
Calibrator
Supply
error (kHz)
kHz
2.500 0 0.017 5
10.000 0 0.07
20.000 0 0.14
30.000 0 0.21
50.000 0 0.35

Procedures: CJ input

1. Connect the instrument to the calibration equipment
(Figure 9).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a one-point
Permitted
DPI 880 error
(V)
0.001
0.002
0.002
0.002
0.002
Gate time = one second
Output = 10V, unipolar, square wave
Frequency = 990 Hz
Input units = Hz (Table 4)
Input trigger level = 5V (Table 4)
Gate time = one second
Output = 10V, unipolar, square wave
Input trigger level = 5V (Table 4)
Units (Table 4): Hz or kHz as
specified in Table 16/17.
Gate time = one second
Units (Table 5): Hz or kHz as
specified in Table 16/17.
Permitted
DPI 880 error (Hz)
0.002
0.001 4
0.002
0.002 1
0.004
0.003 5
0.006
0.005 8
0.011
0.010 4
Permitted
DPI 880 error (kHz)
0.000 2
0.000 042
0.000 2
0.000 112
0.000 3
0.000 205
0.000 4
0.000 298
0.000 6
0.000 483
K0405 Revision B

Publicidad

Tabla de contenido
loading

Tabla de contenido