Operator's Manual 
 
 
illumipro-10 Installation
Installation of the illumipro-10 and its accessories can be performed after contents have been inspected
and the requirements described in this Operator's Manual have been reviewed. Appendix I includes a
general Checklist for Set-up and Installation of the illumipro-10.
The illumipro-10 should be placed on a sturdy, level surface. Set-up the instrument for use as described
in the table below.
Component
Symbol
Printer
Power
12V -----
Supply
@4.5A
External
Keyboard
illumipro-10 Set-up
Instrument set-up is completed under the SYSTEM Menu. The user will be able to set the time format,
date format and preferred language.
illumipro-10 Performance Verification
Performance verification must be performed after installation and prior to use. Optical verification is
completed according to the instructions provided under the SERVICE Menu.
illumipro-10 Operator's Manual: ENGLISH
Copyright© Meridian Bioscience, Inc.
SN11007, REV. 07/2010
All manuals and user guides at all-guides.com
Installation Instruction
Connect the external printer to the illumipro-10 using the supplied
cabling.
The printer connection is located at the back of the
instrument and is identified by the symbol shown.
Secure the printer connection and connect the power supply cable
provided with the printer packaging to the printer.
Connect the power supply cable provided with the illumipro-10
packaging to the unit. The power supply connection is located at the
back of the instrument. The connection port for the power supply is
identified by the symbol shown. Connect the power supply cable to
the power supply box.
Plug the pronged ends of the illumipro-10 and printer power cords to
an appropriate power receptacle.
Connect the optional, external keyboard to the unit using the supplied
cabling. The External Keyboard connection is located at the back of
the instrument and is identified by the symbol shown.
NOTE: The Keyboard must be installed with the illumipro-10 powered
off.