Appendix III
 

illumipro-10 Error Codes and Troubleshooting

The illumipro-10 will display Error Codes when system failures are detected. Error Codes, descriptions
and the actions recommended for correction are summarized in the table below.
Error
Description
Code
500
Real time clock not working.
501
Non-Volatile RAM (NVRAM) not working.
504
Voltage out of range.
505
Voltage out of range.
506
Voltage out of range.
507
Voltage out of range.
508
Interior temperature is too hot.
509
Block A temperature is too hot.
510
Block B temperature is too hot.
Firmware Error: Heater block is not
502
heating properly.
Firmware
503
(NVRAM) did not initialize properly.
Firmware Error: Test results were not
511
fetched from Non-Volatile RAM (NVRAM)
properly.
Firmware Error: Device parameters were
512
not
fetched
(NVRAM) properly.
Firmware Error: Block status cannot be
513
read properly.
illumipro-10 Operator's Manual: ENGLISH
Copyright© Meridian Bioscience, Inc.
SN11007, REV. 07/2010
All manuals and user guides at all-guides.com
Error:
Non-Volatile
RAM
from
Non-Volatile
RAM
Action Required
Contact Meridian Bioscience Technical
Support at (800) 343-3858 for technical
support or to make arrangements for
service.
Move unit to a cooler environment
Turn unit off and wait for it to cool.
Power-off instrument.
Turn instrument back on and allow POST
Testing to be completed. Contact Meridian
Bioscience Technical Support at (800) 343-
3858 for technical support if problem
persists.