9 DIGITAL PRESSURE GAUGE FEATURES & FUNCTIONS (cont.)
9.5 RESETTING THE GAUGE
The gauge has a gauge reset function which restores the original
factory settings.
1
Press
MENU
2
Press
switch off.
3
To cancel the reset request, press
'CANCL' followed by

10 VERIFYING THE GAUGE CALIBRATION

The calibration of the gauge can be verified in the
field using the Elcometer Adhesion Verification
Unit (AVU), part number T99923924, which
connects to the gauge. The pressure reading on
the Elcometer 506 indicator is compared with the
reading of the gauge pressure on the AVU to
verify the calibration. For further information
contact Elcometer or visit www.elcometer.com
11 SPARES & ACCESSORIES
11.1 DOLLIES
The Elcometer 506 can be used with a range of dollies, also referred to
as test elements or stubs. 14.2, 20 and 50mm diameter dollies are
available to purchase as an optional accessory .
14.2mm:
ideal for testing on small surface areas, for measurements
over 25MPa (3600psi) and suitable for use on some
curved surfaces.
20mm:
suitable for use on a variety of coatings / substrates.
50mm:
Coatings on concrete, cementitious layers and uneven
surfaces can be tested more effectively with the larger
50mm dolly. Our 50mm dollies are also available in
stainless steel as required for testing in accordance with
DIN 1048 part 2 and BS EN 12636.
c
Please ensure the appropriate dolly skirt is fitted. See Section 5 - 'Attaching the Gauge to the Dolly' and
Section 11.3 - 'Dolly Skirts' on pages en-5 and en-12 for further information.