LOT identification
The LOT number on this product takes the form of a date clock.
You can see the year at the center of the circle. The arrow points
to the month.
5
Safety instructions
WARNING!
This product may be used only by trained medical staff, who
have been shown how to use it according to the instructions for
use.
Check insulation for damage before every application.
Erbe recommends carrying out an electrical continuity test be-
fore each application of these products.
Note: If the cable conductor is broken or the electrical continuity
of the cable is interrupted in any other manner it can cause arc-
ing in the patient return system or in the active circuit; this can
lead to the patient suffering burns or the outbreak of a fire.
If damaged, do not use these products!
Do not use in the presence of combustible or explosive materi-
als!
Never lay this product on the patient or in his/her direct vicinity.
Protect these products from any form of mechanical damage! Do
not throw! Do not use force!
Do not kink cable.