• In case of battery failure, caused by, for example, an excessive temperature condition (which may resol-
ve itself with time) or by a complete battery failure, the E, F and battery symbol will flash. Turn the pump "off"
and contact your healthcare provider or refer the pump to service.
The power connector at the side of the pump offers the option of connecting the pump to an external alarm system or Patient Data Monitoring
System (PDMS).
• The Flocare
Infinity
Nurse Call (corporate code 35752) enables connection to an external alarm system. This external alarm system may be
®
™
with open or closed contact. Follow the instructions for use of the Flocare
• The Flocare
Infinity
PDMS Cable (corporate code 35776) enables connection to an external PDMS system. Follow the instructions for use of
®
™
the Flocare
Infinity
PDMS Cable to connect the pump.
®
™
• For good pump operation, the pump should be kept clean and free of debris like (dried) tube nutrition. It is recommended to clean the pump with
a wet cloth after every finished feeding regimen and specifically in between patient uses (if applicable).
• Detergent and disinfectant cleaning agents and wipes can damage plastic surfaces of medical devices when they are not compatible with the
surface material. Damaged surfaces may compromise the ability to decontaminate medical devices adequately and / or may interfere with device
function and affect their integrity and performance. Ensure cleaning solutions, detergents and disinfectant wipes are compatible with the device.
Look for signs of damage, cracking, colour changes, etc. to the medical device during and after the application. In case of doubt, contact your local
(Nutricia) sales representative.
• Avoid harsh or agressive cleaners/ disinfectants such as acetone, benzene or iodine.
• Do not sterilize the pump by irradiation, EtO, steam or in an autoclave.
• Always unplug the pump prior to cleaning to avoid electric shock hazard.
• On a regular basis thoroughly clean all surfaces of the pump (including the sensors and rotor) with warm soapy water, a 5% bleach solution in
water, a multipurpose disinfectant cleaner or a soft cloth.
• The Flocare
Infinity
+ pump may be rinsed by holding under a stream of warm, clean water. Do not submerge the pump!
®
™
• Always maintain the rollers on the rotor in a clean state to ensure they spin freely.
• The charger normally does not require cleaning. When desired, a dry or slightly damp cloth may be used to clean the outside surface of the
charger. Make sure the charger is disconnected from the wall outlet.
• Pump disposal: At the end of their service lifetime , the pump and its electrical accessories (AC Adapter Charger and Data Download Accessory
Cable), should be disposed of according to local standards and regulations governing the disposal of electronic waste (e-waste). Other accessories
can be disposed of or recycled as standard non-hazardous waste.
• Strangulation hazard: avoid leaving power adapter cord, feeding set tubing or other choking hazards where infants or young children can become
caught. If these objects get wrapped about one's neck, strangulation can occur.
• The manufacturer recommends an inspection of the pump at an authorised service centre every 2 years.
• If any fault occurs during use, or if the pump is dropped, it should be checked by authorised technical personnel prior to use.
• Replace the Flocare
Infinity
®
Flocare
Infinity
disposable sets properly, as required by local law.
®
™
• Do not use the Infinity alarm systems to trigger actions related to secondary (electrical) medical devices (e.g. a volumetric or syringe pump).
• Do not use pump functions (e.g. the fill set function) for any other purpose than described in this manual, as this may cause the pump to incorrectly
calibrate.
ZMNL900907_AW.indd 25
ZMNL900907_AW.indd 25
CONNECTION TO AN EXTERNAL ALARM SYSTEM
MAINTENANCE AND CLEANING
WARNINGS AND CAUTIONS
pump set every 24 hours to maintain delivery accuracy and prevent the growth of harmful bacteria. Dispose of
™
Infinity
Nurse Call to connect the pump.
®
™
-23-