Condition
f
s
BATT
the pump
remains working
Battery, "E" and "F"
flashing
t
FILL SET
OCC IN
OCC OUT
OCC OUT
(repeated)
LOCK
ER01 - ER99
ER40
No plug symbol visible,
while the pump is con-
nected to the mains.
ZMNL900907_AW.indd 27
ZMNL900907_AW.indd 27
Cause
• The air sensor area is dirty( Fig 1, position 8).
• The feeding set is not inserted correctly.
• The battery capacity is too low.
The pump is not able to deliver
highly viscous fluids at high flow rates with the
present charge level
of the battery.
• Battery failure.
• The pump is priming the set.
• The pump detected an upstream occlusion
between the pump and the feeding bag.
• The pressure sensor area is dirty ( Fig 1, position 5
& 6).
• The pump detected a
downstream occlusion between the pump and
the patient.
• The pressure sensor area is dirty (Fig 1, position 5
& 6).
• Calibration not yet completed
on the current feeding set
• Only the current feeding
schedule is allowed for this patient. Another fee-
ding program is not allowed by your healthcare
professional.
• The self test detected
an electronic error
• Keypad pressed or stuck during pump start up.
• The wall outlet doesn't work.
• The charger is damaged.
Correction
• Make sure the air sensor is clean.
• Make sure the feeding set is properly inserted in the pump. (Fig 5a - 5d)
• Connect the charger to the pump and mains and charge the pump for approx. 6
hours. During charging the pump can be used.
• Turn the pump "off", contact your healthcare provider or refer the pump to
service.
• Press the "FILL SET" key another time to stop the pump and bring it back in the
hold mode.
• Stop the alarm by pressing the "START/STOP" key.
• Remove the feeding set out of the pump and check the permeability by flushing
the line.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the pump and restart the pump.
• Stop the alarm by pressing the "START/STOP" key.
• Remove the feeding set out of the pump and check the permeability by flushing
the line.
• Check the permeability of the feeding tube.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the pump and restart the pump.
• Stop the alarm by pressing the "start/stop" key.
• Start the pump by pressing the "start/stop" key and let it run only briefly.
• Stop the pump by pressing the "start/stop" key insuring that there has been no
occlusion out alarm.
• Remove the feeding set from the pump and reinsert the feeding set in the
pump.
• Restart the pump by pressing the "start/stop" key.
• The programming feature is blocked in the set up mode of the pump. Ask your
healthcare professional to modify this setting.
• Turn the pump "off", make sure the pump door is closed and switch the pump
back "on". If the error persists, contact your healthcare provider or refer the
pump to service.
• Restart pump by ensuring keypad is not pressed.
• Connect the pump to another wall outlet.
• Contact your healthcare professional or
Nutricia subsidiary to replace the charger.