APPENDIX A: GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC
Electromagnetic emissions may affect the operation of any electronic medical de-vice, including enteral feeding pumps.
The Flocare
Infinity
Pump will not be affected by electromagnetic emissions in most environments. However, some electromagnetic fields produced by personal com-
®
™
munication equipment, household appliances, or occupational tools may cause electromagnetic interference (EMI) which can affect the pump.
Possible sources of electromagnetic interference with electronic medical devices include, but are not limited to: cellular phones, cordless telephones, microwave ovens,
anti-theft/security systems, blenders, and high-powered tools (i e drills, saws, chain saws). If electromagnetic emitting devices are operated within one yard/meter of
the Flocare
Infinity
Pump, the pump may automatically shut off and settings may return to their default values. Check the pump regularly if operating near sources of
®
™
electromagnetic emissions.
The Flocare
Infinity
Pump can safely be operated on commercial aircraft and is designed in accordance with EN 60601-1-2, EN 60601-1-4 and RT CA DO160D
®
™
standards for electromagnetic emissions and immunity.
Guidance and manufacturer's Declaration – Electromagnetic emissions
The Flocare
Infinity
enteral feeding pump is intended for use in the electromagnetic environment specified below.
®
™
The customer or the user of the Flocare
environment.
Emissions test
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Guidance and manufacturer's Declaration – Electromagnetic immunity
The Flocare
Infinity
enteral feeding pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Flocare
®
™
enteral feeding pump should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electric fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interrupti-
ons and voltage variations on
power supply lines.
IEC 61000-4-11
ZMNL900907_AW.indd 31
ZMNL900907_AW.indd 31
COMPATIBILITY
Infinity
enteral feeding pump should assure that it is used in such an
®
™
Compliance
Group 1
Class B
Class A
Complies
IEC 60601 test level
Compliance level
± 8 kV contact
± 8 kV contact
± 15 kV air
± 15 kV air
+/- 2 kV @ PRR
+/- 2 kV @ PRR
100kHz for power supply lines
100kHz for power supply lines
± 1 kV @ PRR
± 1 kV @ PRR
100kHz for input/ output lines
100kHz for input/ output lines
± 1 kV line to line
± 1 kV line to line
70 % UT; 25/30 cycles single
70 % UT; 25/30
phase:at 0°
cycles single phase:
0 % UT; 1 cycle at 0°
at 0°
0 % UT; 0,5 cycle at
0 % UT; 1 cycle at 0°
0°, 45°, 90°, 135°,
0 % UT; 0,5 cycle at
180°, 225°, 270° and
0°, 45°, 90°, 135°,
315°
180°, 225°, 270° and 315°
0 % UT; 250/300
0 % UT; 250/300
cycle (5s)
cycle (5s)
Electromagnetic environment – guidance
The Flocare
Infinity
uses RF energy only for its internal function. Therefore, its
®
™
RF emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
The Flocare
Infinity
is suitable for use in all
®
™
establishments, including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies buildings used for
domestic purposes.
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. The Flocare
operation during power mains interruptions via the internal
battery.