• No modification of this equipment is allowed.
• The AIR alarm should not be used to indicate an end of dose. NOTE: do not mute the audio alarm for end of dose alarm in these instances ('end of
dose' will appear in the pump screen).
• Non-liquid food ingredients and mixed or reconstituted foods may have a tendency to occlude flow within the feeding tube and feeding set. They
may also have the tendency to adhere to the tubing wall and obstruct the enteral feeding pump optical air-in-line sensor. Clinicians and caregivers
should carefully monitor the pump during use with these type of feedings.
• The Flocare
Infinity
+ pump delivers the dose within +/-5% volumetric accuracy under the following conditions:
®
- utilizing Flocare Infinity disposable giving sets
- fluid head height at +152 mm ± 76 mm with respect to center of rotor.

ALARM FUNCTIONS AND SAFETY FEATURES

In case any of the problems listed on the problem solver chart occur, the pump delivers an audible and visual alarm and stops working. The back light
of the LCD automatically switches on.
Exception to this is the low battery alarm "BATT", in this situation the pump continues working.
Action in the event of an alarm:
Check the type of alarm displayed by the large display characters.
• Press the "start/stop" key to stop the audible and visual alarm.
• Correct the cause of the alarm as described in the table.
• Start the pump again by pressing "start/stop".
The problem solver chart gives a clear explanation of the alarms.
Using methods other than those described may cause the pump to function incorrectly.
Condition
NO SET
• The set is not fitted or wrongly fitted in the pump.
• The pressure sensor area is dirty
PUSH STRT
• The pump has been untouched in hold mode for 3
END OF DOSE
• The pump administered the installed dose =
PROG
• No flow rate is installed.
• The installed parameters in a bolus program can-
DOOR
• The door is not correctly closed.
AIR
• The pump detected air in the pump set.
ZMNL900907_AW.indd 26
ZMNL900907_AW.indd 26
UK

PROBLEM SOLVER CHART

Always follow the instructions below in case a problem occurs.
Cause
(Fig 1, position 5 & 6)
minutes or more.
volume.
Flow rate = 0 ml/h
not form a correct
solution.
Correction
• Stop the alarm by pressing "start/stop".
• Insert the Flocare
Infinity
feeding set into the pump as indicated on the blister
®
packaging (or refer to Fig 5a - 5d) and close the door.
• Restart the pump.
• Clean the sensors, reinsert the feeding set in the pump and restart the pump.
• Stop the alarm and prolong the hold mode with another 3 minutes by pressing
"start/stop" key.
• Program the pump and start it by pressing the "start/stop".
• The programmed dose/volume has been administered
• Press start/stop to switch off the alarm
• Press INFO and then CLR to reset the administered volume (to reset the flow
rate and dose as well, press the CLR button for 3 seconds)
• Make sure the correct feeding parameters are programmed.
• Make sure the door is properly closed prior to
starting a feeding program.
• Make sure there is no excessive foam in the reservoir and ensure there is
enough feed in the reservoir for the set therapy.
• If the feeding reservoir is empty, replace the reservoir and continue feeding. If
necessary, prime the set (FILL SET).
-24-
12.05.21 15:59
12.05.21 15:59