Guidance and manufacturer's Declaration – Electromagnetic immunity
The Flocare
Infinity
enteral feeding pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Flocare
®
™
enteral feeding pump should assure that it is used in such an environment.
Immunity test
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Recommended separation distances between portable and mobile RF communications equipment and the
The Flocare
Infinity
enteral feeding pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or
®
™
the user of the Flocare
Infinity
®
™
equipment (transmitters) and the Flocare
Rated Maximum output power of
transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation appli-
cable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in Watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
ZMNL900907_AW.indd 32
ZMNL900907_AW.indd 32
IEC 60601 test level
Compliance level
30 A/m
30 A/m
Flocare
Infinity
®
can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
Infinity
as recommended below, according to the maximum output power of the communications equipment.
®
™
Separation distance according to frequency of transmitter (m)
150 KHz to 80 MHz
d = 1.2
CD
P
0.12
0.38
1.2
3.8
12
Electromagnetic environment – guidance
Power frequency magnetic fields should be at levels characte-
ristic of a typical location in the typical commercial or hospital
environment.
enteral feeding pump
™
80 MHz to 800 MHz
d = 1.2
CD
P
0.12
0.38
1.2
3.8
12
-31-
Infinity
®
™
800 MHz to 6 GHz
d = 2.3
CD
P
0.23
0.73
2.3
7.3
23