Oxygen Analyzer
cloth moistened with ethanol or Cidex. The instrument
is not designed to withstand the conditions imposed by
steam, ethylene oxide or radiation sterilization.
Sensor Deflector, Retaining Strap and Tee
Adapter
The sensor deflector, retaining strap and tee adapter
may be cleaned by wiping them with a cloth moistened
with a mild detergent solution. The parts must be
thoroughly dry before they are used.
Disinfection
The sensor deflector, retaining strap and tee adapter
may be disinfected by washing them with ethanol or
Cidex (per manufacturer's instructions). The parts must
be thoroughly dry before they are used.
5-2
All manuals and user guides at all-guides.com
Sterilization
The sensor deflector, retaining strap and tee adapter
may be sterilized using Cidex (per manufacturer's
instructions), steam or ethylene oxide. Due to the
varying conditions imposed on materials during
sterilization, it is not possible to determine the exact
number of times sterilization processes can be carried
out. Therefore, the operator must carefully examine the
sensor deflector, retaining strap and tee adapter after
sterilization and prior to use to verify that the items are
fit for use. The operator must verify that the items are
free from tears and cracks and that the items have not
undergone any material changes that may compromise
their fitness for use (e.g., brittleness and dimensional
changes). The operator must also examine the items
to verify that the items are free of chemical residuals
resulting from the sterilization process.
Because of the variability of cleaning, disinfection
and sterilization processes, Ohio Medical Corporation
cannot provide specific sterilization instructions, nor can
the sterility of an item be ensured
P/N 711462 (Rev.3) 02/2016
Section 5, Maintenance and Care