Installation
RELAY connection
Connection side view
USB type B
USB type B
4
3
1
2
8
The switch functions and the error monitoring influence the position of various relays.
The relay contacts can be used for switching via the RELAY terminal. The relay contacts
are isolated.
Dangerous contact voltage
Voltages above 30 V AC or 60 V DC are deemed to be dangerous in the event of contact
in accordance with EN 61010.
Only apply a grounded protective extra-low voltage (PELV).
Connect peripheral components with a shielded connecting cable (electromagnetic
compatibility) to the RELAY terminal on the back of the device.
Pin assignment and contact positions
The pins on the 25-pin sub-D socket are assigned as follows:
Pin
Signal
Switch function 1
4
5
6
Switch function 2
8
9
10
Switch function 3
11
12
13
Switch function 4
16
17
18
Switch function 5
19
20
21
Switch function 6
22
23
24
Error signal
3
15
14
Infeed for relays with a higher switching capacity
25
+24 V DC, 200 mA
1, 7
GND
2
n.c.
1)
Fused with a PTC element at 200 mA, self-resetting after the device power-off or disconnecting of
the RELAY plug. Meets the requirements for grounded protective extra-low voltage.
The B-type USB port facilitates direct communication with the TPG 366 via a computer
(e.g. firmware update, parameter storage (read/write)).
Pin
Assignment
1
VBUS (+5 V)
2
D-
3
D+
DANGER
Pressure higher than
threshold or device
switched off
Pressure higher than
threshold or device
switched off
Pressure higher than
threshold or device
switched off
Pressure higher than
threshold or device
switched off
Pressure higher than
threshold or device
switched off
Pressure higher than
threshold or device
switched off
Error or device switched off
1)
Pressure lower than
threshold
Pressure lower than
threshold
Pressure lower than
threshold
Pressure lower than
threshold
Pressure lower than
threshold
Pressure lower than
threshold
No error