7. Using a Mains Adapter

You can operate this instrument using the Microlife mains adapter
(DC 6V, 600mA).
Only use the Microlife mains adapter available as an orig-
inal accessory appropriate for your supply voltage, e.g. the
«Microlife 230V adapter».
Ensure that neither the mains adapter nor the cable are
damaged.
1. Plug the adapter cable into the mains adapter socket 4 in the
blood pressure monitor.
2. Plug the adapter plug into the wall socket.
When the mains adapter is connected, no battery current is consumed.

8. Error Messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
Error
Description Potential cause and remedy
«ERR 1» Signal too
The pulse signals on the cuff are too
weak
weak. Re-position the cuff and repeat the
measurement.*
«ERR 2» Error signal During the measurement, error signals
were detected by the cuff, caused for
instance by movement or muscle tension.
Repeat the measurement, keeping your
arm still.
«ERR 3» No pres-
An adequate pressure cannot be gener-
sure in the
ated in the cuff. A leak may have
cuff
occurred. Check that the cuff is correctly
connected and is not too loose. Replace
the batteries if necessary. Repeat the
measurement.
«ERR 5» Abnormal
The measuring signals are inaccurate
result
and no result can therefore be displayed.
Read through the checklist for performing
reliable measurements and then repeat
the measurement.*
BP A150
Error
Description Potential cause and remedy
«ERR 6» MAM Mode There were too many errors during the
measurement in MAM mode, making it
impossible to obtain a final result. Read
through the checklist for performing reli-
able measurements and then repeat the
measurement.*
«HI»
Pulse or cuff
The pressure in the cuff is too high (over
pressure too
300 mmHg) OR the pulse is too high (over
high
200 beats per minute). Relax for 5
minutes and repeat the measurement.*
«LO»
Pulse too
The pulse is too low (less than 40 beats
low
per minute). Repeat the measurement.*
* Please consult your doctor, if this or any other problem occurs
repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.

9. Safety, Care, Accuracy Test and Disposal

Safety and protection

 This instrument may only be used for the purposes described in
this booklet. The manufacturer cannot be held liable for damage
caused by incorrect application.
 This instrument comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical Specifications» section!
 Protect it from:
- water and moisture
- extreme temperatures
- impact and dropping
- contamination and dust
- direct sunlight
- heat and cold
 The cuffs are sensitive and must be handled with care.
 Only pump up the cuff once fitted.
 Do not use the instrument close to strong electromagnetic fields
such as mobile telephones or radio installations.
 Do not use the instrument if you think it is damaged or notice
anything unusual.
 Never open the instrument.
EN