Emc Table - i-SENS KetoSens BT Manual Del Usuario

Sistema de monitoreo de b-cetona en sangre
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

EMC Table

The following tables contain the Manufacturer's declaration and
additional information required by IEC 60601-1-2:2014 (Fourth
Edition).
Phenomenon
Radiated disturbance
Electrostatic
Discharges (ESD)
Radiated RF
Electromagnetic
Fields
Electric Fast
Transients and bursts
Surges
Conducted
Disturbances,
induced by RF fields
Power-frequency
Magnetic Field
Basic EMC standard
or test method
CISPR 11: 2015 +A1:
2016
IEC 61000-4-2:2008
IEC 61000-4-3:2006
+ A1:2007+A2:2010
IEC 61000-4-4:2012
IEC 61000-4-5:2014
IEC 61000-4-6:2013
IEC 61000-4-8:2009
Port tested
requirement
Group 1,
Enclosure
Class B
±8 kV
Contact
Enclosure
±2 kV, ±4 kV,
±8 kV, ±15 kV Air
3 V/m
80 MHz to 2.7 GHz
80%, 1 Hz AM
Enclosure
RF Wireless Comm.
(Refer to test report
clause 1.15)
±2 kV AC,
AC Mains
100 kHz PRR
±0.5 kV, ±1 kV
AC Mains
L1 to L2 (DM)
3 Vrms
150 kHz to 80 MHz
6 Vrms in ISM
AC Mains
bands between
0.15 MHz and
80 MHz
Enclosure
30 A/m
www.i-sens.com 57