WARNING
Risk of deviating measured values during flow
measurement.
Applies to Narkomed 6000:
Manual ventilation may result in deviating measured
values.
Replacing/Snapping into place
The disposable CO
absorber must be replaced as soon as one of
2
the following conditions applies:
–
The inspiratory CO
concentration has reached approx.
2
5 mmHg.
–
The soda lime has turned 2/3 violet.
–
The disposable CO
absorber has been used for 4 weeks or
2
longer.
NOTE
The color change is reversible. Therefore, depleted soda lime
can be accidentally considered to be unused. Regarding when to
replace the soda lime, the color change is only a rough indicator.
The CO
monitoring is the decisive factor.
2
Always remove depleted disposable CO
device and dispose of immediately.
Replace the disposable CO
2
continuous absorption of CO
2
1
Press the button: the disposable CO
2
Remove the disposable CO
3
Before fitting into place, shake the new disposable CO
absorber by turning it upside down several times in order to
loosen up the soda lime.
Make sure to fit the disposable CO
correctly (pictograms must face the user).
4
Remove the red cap.
5
Slide the disposable CO
absorber onto the mounting as far as
2
possible.
16
absorbers from the
2
absorber quickly to ensure
in the breathing system.
absorber swings out.
2
absorber from the mounting.
2
2
absorber into place
2
Instructions for use CLIC Absorber 800+ / Infinity ID CLIC Absorber 800+ / CLIC Adapter
6
Swing into the system until it snaps into place.
The disposable CO
absorber must be fully covered by the
2
CLIC adapter.
CO
bypass function
2
WARNING
Risk of patient injury
If the absorber is swung out, there is no absorption of CO2
in the breathing system.
Activating the CO
bypass function
2
Press the button: the disposable CO
Deactivating the CO
bypass function:
2
Swing the disposable CO
snaps into place.
absorber swings out.
2
absorber into the device until it
2