Fig. 13
The device drains by gravity and therefore, the drain pipe must descend with respect to the
machine.
Prevent any upward bend in the drain pipe (siphons) (Fig. 14) that could generate
stagnation. There must be no non-return valves.
ATTENTION: The sewage drain must be lower than the drain outlet of the machine
the connection pipe must not be bent below the sewage drain.
In the absence of a sewer line, in a ventilated area, you can use a tank
accumulation:
- Do not immerse the exhaust chamber in the liquid accumulated in the tank;
- The tank should not be sent to pressure or sealed.
6.2
Filling the disinfectant/detergent and lubricant tanks:
The tanks can contain over 1 l;
however, do not exceed this amount when filling the tanks.
ALWAYS SWITCH THE MACHINE ON BEFORE FILLING THE TANKS
Fig. 14
After connecting the above, move the main switch behind the device (Fig.
15) from 0 to I;
The display lights up and the manufacturer logo appears for a few seconds
(screen A) and immediately after message indicating a low level in the tank
(Para. 9: Screen A1).
Fig. 15
ENGLISH
(Fig. 14)
and
13