Errors: Causes And Solutions - Tecno-gaz Ekstraplus PT001TTM Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

ENGLISH
7.7

Errors: causes and solutions

Situations that may occur prior to operation and prevent the cycle from being launched:
EMPTY DETERGENT TANK
FLASHES
SCREEN E101
FULL DETERGENT TANK
FLASHES
SCREEN E102
EMPTY OIL TANK
FLASHES
SCREEN E103
FULL OIL TANK
FLASHES
SCREEN E104
SD NOT INSERTED
CYCLE NOT SAVED
SCREEN E105
24
OK
OK
OK
OK
OK
CAUSES:
This is displayed when the user presses
START and the level of detergent is insufficient
to perform a cycle.
SOLUTIONS:
Pour detergent into the tank.
CAUSES:
This is displayed when the level reaches the
maximum
limit
while
accompanied by a beep.
SOLUTIONS:
Suspend the filling operations and put the cap
on the filler inlet of the tank.
CAUSES:
This is displayed when the user presses
START and the level of oil is insufficient to
perform a cycle.
SOLUTIONS:
Pour oil into the tank.
CAUSES:
This is displayed when the level reaches the
maximum
limit
while
accompanied by a beep.
SOLUTIONS:
Suspend the filling operations and put the cap
on the filler inlet of the tank.
CAUSES:
Screen E105 appears if START is pressed and
no SD is inserted.
SOLUTIONS:
The cycle is launched and not saved if OK is
pressed to confirm. Insert the SD for the cycle
to be launched and saved.
filling
the
tank,
filling
the
tank,

Publicidad

Tabla de contenido
loading

Tabla de contenido