40 | Ottobock
Plugging in the contacts
Lubricate the cable connectors with 633F11 Silicone
Grease.
2
Plug in the cables, being certain they are fully inserted
into the connectors.
3
Protect from moisture damage by applying 633F11
Silicone Grease liberally to the top of connectors and
cover.
4
Mounting the housing
Push the electrode housing, including the installed
wires, into the lubricated lamination ring and press
firmly. It must completely lock in!
NOTICE
Misalignment during installation. Misalignment
during installation creates the risk of loss of contact
5
or short-circuiting of the spring contacts. The me-
chanical strength can also be reduced due to thread
misalignment.
The illustration shows an electrode housing that is not
locked in properly, it is askew.
6
DynamicArm