Troubleshooting
If the motor is not running
Check if:
– the Clario is switched on�
– the mains voltage is correct and if the mains plug is
inserted correctly into the wall socket�
– the internal battery is charged (only AC / DC version
under battery operation)�
If the pump is too loud
Check the position and seal of the valve discs inside the
SafetyChamber� If necessary reposition according to
chapter "Replacing the valve discs"�
If the pump does not suction strongly enough
Check if:
– the vacuum regulator is set to the correct and
prescribed suction level�
– the tubings are not defective or broken� If necessary,
replace�
– all plug-in connections are tight�
– the suction canister / lid have no cracks, brittle areas,
discolouration� If necessary, replace�
– the seal of the valve discs inside the SafetyChamber
is tight�
– the membrane is clean and has no tears�