Battery calibration

(AC / DC version only)
CAUTIONS
Calibrate the rechargeable battery before first use
and after storage duration > 2 months� Replace
battery if the calibration is not successful�
Calibrating NiMH rechargeable battery
1�
Make sure that the rechargeable battery is fully charged�
2�
Pull out the mains plug from the fixed mains socket�
3�
4�
Let the pump run until the rechargeable battery is
empty and the pump comes to a stop�
5�
Connect the Clario to a fixed mains socket� The
rechargeable battery is being charged�
6�
Wait until the charging indicator goes out�
7�
Repeat step 1 to 6� The Clario AC / DC is then ready
to use�
Storage of the rechargeable battery
In order to keep the self-discharge at a minimum, store
the Clario and the removed replacement rechargeable
batteries at temperatures below 25 °C (77 °F)� Repeat the
calibration process every 60–90 days�
24
Switch on the Clario�
The pump will start running�