Testing the rechargeable battery
1�
Make sure that the rechargeable battery is fully
charged�
2�
Pull out the mains plug from the fixed mains socket�
3�
4�
5� Measure the time until the pump comes to a stop�
The battery is ok, when the running time is
> 50 minutes� If the running time is < 50 minutes,
repeat test or calibrate the battery (see next page)�
Set vacuum regulator to position
"max�"�
Switch on the Clario�
The pump will start running�