Connection to an APC 2:
WARNING! Risk of injury! Do not activate the electrosurgical unit
while you are connecting the adapter.
1. Switch on the VIO electrosurgical unit and select the basic
program.
2
2. Insert the HF connector of the adapter (1) into the multifunc-
tional receptacle of the APC 2. The connector must distinctly
click.
3. Screw the argon gas connection of the adapter (2) into the
argon gas receptacle of the APC 2.
4. When the adapter is connected, the standard settings for
mode, power, effect, and gas flow are configured automati-
cally on the VIO electrosurgical unit. After connecting an APC
probe to the adapter, the standard settings may have to be
adjusted for the particular APC probe that is being used. It is
essential that you read the recommendations provided in the
instructions for that APC probe.
4.4
After use
The adapter can remain in the unit after use and be cleaned and
disinfected together with it. In this case, please read the pro-
cessing recommendations in the User Manual for the relevant
device.