REMOVAL OF EXCESS AIR FROM THE HYDRAULIC SYSTEM

Tools Required:
• None
Procedure:
1 .
Verify that all hydraulic linkages are secure and operating properly .
2 .
Raise the litter to its highest height .
3 .
Continue to pump the pedal several times to force the air through the system .
4 .
Verify proper operation before returning the unit to service .