Citizen CH-456 Manual De Instrucciones página 20

Ocultar thumbs Ver también para CH-456:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 26
ELECTOROMAGNETIC COMPATIBILITY INFORMATION
WARNING
• Portable RF communications equipment should be used no closer than 30 cm (12
inches) to any part of the [CH-456], including cables specified.
• Use of this equipment adjacent to or stacked with other equipment should be
avoided.
• Use of accessories and options other than those specified (other than CITIZEN
original parts) could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment.
Guidance and manufacturer's declaration - electromagnetic emissions
The [CH-456] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [CH-456] should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The [CH-456] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [CH-456] should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Electrostatic
±8 kV contact
discharge (ESD)
IEC 61000-4-2
±15 kV air
Electrical fast
±2 kV for
transient/burst
power supply
IEC 61000-4-4
lines
±1 kV for
input/output
lines
Surge
±1 kV
IEC 61000-4-5
line to line
±2 kV
line to earth
Voltage dips,
0% U
short
0.5 cycle
interruptions
0% U
and voltage
1 cycle
variations on
70% U
power supply
25/30 cycle
IEC 61000-4-11
0% U
250/300 cycle
Power frequency
(50/60 Hz)
30 A/m
magnetic field
IEC 61000-4-8
Note: U
is the A.C. mains voltage prior to application of the test level.
T
Compliance
Group 1
Class B
Class A
Complies
Compliance
level
±8 kV contact
±15 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV
line to line
±2 kV
line to earth
0% U
T
T
0.5 cycle
0% U
T
T
1 cycle
70% U
T
T
25/30 cycle
0% U
T
T
250/300 cycle
30 A/m
- Eng 19 -
Electromagnetic environment - guidance
The [CH-456] uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
The [CH-456] is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
[CH-456] requires continued operation
during power mains interruptions, it is
recommended that the [CH-456] be
powered from an uninterruptible power
supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.

Publicidad

Tabla de contenido
loading

Tabla de contenido