TECHNICAL FEATURES
DEVICE
Size:
Weight:
Case:
Protection level:
Storage and transportation temperature:
Max. relative humidity:
The values indicate the limits allowed if the product or its accessories are not in the original
package.
USE CONDITIONS
Temperature:
Max. relative humidity:
Atmospheric pressure:
TECHNICAL FEATURES OF THE CURRENTS
EMS and TENS
Channels available:
Constant current:
Intensity:
Wave form:
Working frequency:
Recovery frequency:
Pulse amplitude:
Working time:
Recovery time:
Frequency modulation range:
Min. modulation time:
Amplitude modulation range:
MICROCURRENTS
Channels available:
Constant current:
Min. frequency:
Max. frequency:
Min. Intensity:
Max. Intensity:
Amplitude value:
DENERVATED MUSCLES
Channels available:
Max. Intensity:
Pulses:
IONOPHORESIS
Channels available:
Constant current:
Min. intensity:
Max. intensity:
Min. time:
Max. time:
36
160 x 99 x 35,4 mm
404 g.
in food grade ABS
IP 22
from -10°C to 45°C
30% - 75%
from 0°C to 35°C
from 15% to 93%
from 700 hPa to 1060 hPa
Channels 1-2-3-4
Yes
0-120 mA with 1000 Ohm load
rectangular, biphasic, symmetric, compensated
0,3-150 Hz
0,3-150 Hz
30-1000 µs
from 1 to 30 seconds
from 0 to 1 minutes
cont. variation from 1 to 150 Hz
3 seconds
continuous variation from 50 to 450 seconds
Channels 1-3
Yes
5Hz
200Hz
0 µA/1000 Ohm steps of 10 µA
800 µA/1000 Ohm
Included between 1 and 250 µseconds
Channels 1-3
60 mA
Triangular 1000 ms, Rectangular/trapezoidal 500 ms
Channel 1
Yes
0 mA/1000 Ohm
10 mA/1000 Ohm steps of 0.1 mA/1000 Ohm
1 minute
99 minutes