IndIcatIon

The proGAV 2.0 is used for cerebrospinal fl uid
(cSf) drainage in the treatment of hydrocepha-
lus.

technIcal descrIPtIon

The proGAV 2.0 is a posture dependent hydro-
cephalus valve. It comprises an adjustable dif-
ferential pressure unit and a fi xed gravitational
unit (fi g. 1 and 2).
Fig. 1: proGAV 2.0 - side view
1 sapphire ball
2 bow spring
3 rotor
4 tantalum ball
5 sapphire ball
6 connector under the silicone catheter
Fig. 2: Schematic cross section of the proGAV 2.0
18
The adjustable dP-unit is composed of a solid
titanium body with a well-tried ball-cone valve
(1) integrated in its proximal part. A bow spring
(2) defi nes the opening pressure of the ball-
cone valve. The pretensioning of the spring,
and thus the valve opening pressure, can be
adjusted by turning a rotor (3), with the valve
implanted under the patient's skin.
The gravitational unit contains a tantalum ball
(4), which defi nes the opening pressure of this
valve, and a sapphire ball (5), which ensures the
precise closure of the valve. Underneath the
silicone catheters (6), a connector is provided.
adjustable differential pressure unit
(adjustable DP-unit)
gravitational unit
1
2
3
6
6
4
5