Gima 34244 Manual De Uso Y Mantenimiento página 15

Bolsas autoclavables de silicona
Ocultar thumbs Ver también para 34244:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

15
Product description and intended use
GIMA Manual Resuscitator is designed for use as an adjunct to
artificial respiration and cardiopulmonary resuscitation. The Resuscitator can be used to ventilate the
apnoeic patient and to augment ventilation and for oxygen delivery to the spontaneously breathing
patient. The designs are also different according to Adult, Child or Infant by using different compressing
frequency, they also come with different sizes to meet all patients' need for oxygen.
The adult size has a Pressure Relief Valve design (as an option). When the pressure inside the bag
exceeds 60 cmH
0, and 40 cmH
2
delivered breath tn the atmosphere to protect the lung from injuring by the high pressure.
This product is intended for use by qualified medical or emergency personnel trained in pulmonary
ventilation and advanced cardiac life support techniques. This manual provides all section assembly
drawings, numbers for all construction parts. some cautions. and cleaning methods. Please carefully read
all cautions in this manual before use to accomplish the best effect, and pay attention to all safety warnings.
GIMA Silicone Resuscitator
It is manufactured by using supreme rated Silicone, with high flexibility, stable material, can resist high
temperature (to a maximum of 134°C).
Warning & Caution
WARNINGS
1. Do not use the Manual Resuscitator in toxic atmospheres.
2. Remove the oxygen reservoir and reservoir valve if supplemental oxygen is not being
administered. Failure to do so will affect the refill rate and maximum frequency capabilities.
3. Do not administer supplemental oxygen in the presence of open flames.
4. Do not use oil, grease or other hydrocarbon-based substances on any part of the manual
Resuscitator Supplemental oxygen, supplied under pressure, can combine with hydrocarbons
and cause explosions.
5. This device is intended for use by qualified medical and emergency
personnel trained in pulmonary ventilation and advanced cardiac life support techniques.
6. Proficiency in the assembly, disassembly and use of this device should be demonstrated
before use on a patient.
7. Always test the device in accordance with this manual after cleaning and sterilization or
replacement of parts.
8. Always monitor airway pressure with a manometer when ventilating a patient.
9. Only qualified personnel trained in the use of positive end expiratory pressure (PEEP)
should administer PEEP with this device.
10. Always verify PEEP level and the function of the resuscitator before
use on a patient.
11. Monitor airway pressure with a manometer when ventilating a patient.
CAUTIONS
1. If overriding the pressure relief valve, great caution must be taken not to allow the pressure in the
patients airways became too high.
2. Do not attempt to disassemble the pressure relief valve assembly. Disassembly will damage the
component.
3. Before use, clean and sterilize the entire Manual Resuscitator to your individual institution's validated
procedure for cleaning and sterilizing such equipment. After the Manual Resuscitator has been cleaned
and sterilized, test the Manual Resuscitator as directed in this manual.
0 for infant bag, the Pressure Relief Valve will automatically vent the
2
ENGLISH

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

342453424634247

Tabla de contenido