Sterilization methods
Note for sterilizing timing: When first use of the new resuscitator and when user changed do
sterilization for the resuscitator. Also when the resuscitator was used for more than 48 hours
do sterilization for the resuscitator.
Autoclaved sterilization max (134°C): Can be used on all parts of the resuscitator, except PVC material
products. Follow the recommendations given by the steam autoclaving manufacturer. All steam
autoclave cycles used for porous items are acceptable provided that the maximum temperature does
not exceed 273°F or 134°C. If plastic housing made by polycarbonate the maximum temperature
does not exceed 121°C (244°F).
Ethylene oxide sterilization: Can be used on all parts of the resuscitator. Follow ethylene oxide
chamber manufacturer's instruction.
Gamma radiation sterilization: Can be used on all parts of the resuscitator. Follow gamma radiation
sterilization manufacturer's instruction.
Visual inspection through package assembly
1. After cleaning-inspection-sterilization process, let the resuscitator parts stay at room temperature to
dry (air dry). Wait until all parts are dry. No drying agent is needed.
2. After cleaning-disinfection-sterilization carefully inspect all parts for damage or excessive wear and
replace them if necessary. Some methods may cause discolouration of rubber parts but will not affect
their lifetime. In case of material deterioration, eg. cracking. the parts should be replaced. Contact your
distributor for part replacement.
3. Assemble the parts following the Assembly View.
4. After assembling the resuscitator, a label should be put on to indicate the handling date.
Storage
Far compact storage, e.g. in an emergency case, the inlet end can be pushed halfway into the bag.
Never store the resuscitator in a compressed folded state.
Never excessively squeeze the bag during storage. When the resuscitator is ready for use it should
not be kept in direct sunlight or in a heated environment.
Storage temperature: -40°C to 70°C (-40°F to 158°F)
For long-term storage or transportation the resuscitator should be kept in closed packing in a cool
place away from direct sunlight.
Applicable
Methods
Parts
A: Non-rebreathing valve
B: Resuscitator body
C: Reservoir valve
D: Oxygen reservoir
E: Oxygen tubing
F: Mask
: applicable
: not applicable
Cleaning
(Washing)
Manual
Washing
washing
machine
disinfecting
(WM)
Silicone:
Silicone:
Silicone:
PVC:
PVC:
PVC:
Silicone:
Silicone:
Silicone:
PVC:
PVC:
PVC:
ENGLISH
Disinfecting
WM
Boiling
Chemicals
Silicone:
Silicone:
PVC:
PVC:
Silicone:
Silicone:
PVC:
PVC:
Autoclaving
121°C (244°F)
134°C (273°F)
Polycarbonate
Polysulfone
Polycarbonate
Polysulfone
Polycarbonate
Polysulfone
Silicone:
Silicone:
PVC:
PVC:
Silicone:
Silicone:
PVC:
PVC: