After the recharging process is completed, you should always remove the mains plug from the
socket first and then remove the connector plug from the control box.
Do not use the lift while it is being recharged!
If the batteries are not used for a significant period, they discharge slowly by themselves (deep
discharge). It then becomes impossible to recharge them. You should therefore recharge the
batteries at least once a month even if they are not being used.
3.4

Emergency

In cases for emergency the emergency stop and the function for lowering in an emergency must
be used.

3.4.1 Emergency stop

In emergencies, and during assembly and disassembly, the
emergency stop is activated by pressing the red button on the
control box.
Push on the button and turn in the direction of the arrows. If the
button is fully pressed to the back turn back in the opposite
direction of the arrows.
The emergency stop is deactivated by turning the red button in
the direction of the arrows.

3.4.2 Emergency lowering

WARNING: Risk of injury – Adjust the manual emergency lowering for the weight of
L
the patient.
Emergency lowering if the power fails or the batteries are empty
can be done using the red pull button  at the lower end of the
motor. This emergency lowering is factory-set for a patient weight
of 75 kg.
Standard calibration: 3000N, 16 mm/s to lower the patiënt.
Please note that the manual emergency lowering is only possible
when the patient is sitting in the patient hoist.