• Setup for Data Logger parameter
In measure mode press button
•
Use the directional keys to move to "LOG" menu
•
.
•
Move with the keys
The table below shows the setup menu structure for the Data Logger mode; for each program, there are
the options that the user can choose and the default value:
Program
Data logging type
Delete data in memory
Data logging type
Access this menu to select the data acquisition mode:
MANUAL: The data is acquired only when the user presses the button
•
•
SECONDS – MINUTES - HOURS: Set an automatic data acquisition frequency range.
Use the directional keys to move from MANUAL to HOURS or MINUTES. Access with
directional keys
and change the value of the acquisition time. Confirm the setting with the key
• Use of automatic Data Logger
Press the button
in measure mode, to start and end the automatic recording.
When the automatic data saving is running, the icon
When it is set, but not in operation, the icon on the screen remains fixed.
Note: scrolling the parameters, the recording stops.
IMPORTANT: for recordings lasting longer than 15 hours, it is recommended to connect the device to an
external power source (PC or electrical outlet) through the appropriate cable.
• Achievement of limit memory (10000 total values)
The following reports inform the user of the acheivement of the maximum instrumental data saving
capacity:
•
When the 9500 total data stored is reached, the icon
When the emory limit is reached, the data logger stops automatically and the string "End of Log
•
Memory / Delete Memory" appears next to the error signal
Delete data in memory
Access this menu and select YES to delete the saved data and empty the memory.
• Example of automatic Data Logger mode
Example: automatic pH recording on internal memory every 2 minutes
Access the "LOG" setup menu
•
•
Press the button
UM ReVio EN rev.0 04.06.2020
to access the SETUP menu.
and
to select the program to access.
Description
Manual – Seconds – Minutes - Hours
Yes - No
.
, enter the Data logging type menu and move with directional keys to MINUTES.
41
and access the menu by pressing the key
Options
.
flashes on the display.
starts flashing.
.
Factory Default Settings
Manual
No
and with the
.
EN