2. Important Information
2.1 Disclaimers
Use only steam-process distilled water in your Statim. Deionized, demineralized, or specially
filtered water should not be used. Never use tap water.
Do not permit any person other than certified personnel to supply parts for, service or
maintain your Statim. SciCan shall not be liable for incidental, special or consequential
damages caused by any maintenance or services performed on the Statim by a third party,
or for the use of equipment or parts manufactured by a third party, including lost profits, any
commercial loss, economic loss, or loss arising from personal injury.
Never remove the cover of the unit and never insert objects through holes or openings in the
cabinetry. Doing so may damage the unit and / or pose a hazard to the operator.
All elements of this book are common to both Statim 2000S and Statim 5000S except when
noted.
IMPORTANT:
Follow local guidelines governing verification of the sterilization procedure.
Drying Performance
The Statim 2000S and 5000S have been designed to provide a complete sterilization
solution for your unwrapped and wrapped instrument needs: rapid sterilization balanced with
rapid drying, by using the SciCan Dri-Tec Drying Technology.
The Statim 2000S uses convection heat to dry instruments by utilizing the remaining heat
in the system after the sterilization phase. Heat is captured and released in the cassette to
rapidly dry a properly loaded Statim cassette.
The Statim 5000S uses heat generated by the sterilization phase which is absorbed by
the drying plates. The heat is transferred from drying plates directly to the load resulting in
accelerated, rapid drying of a properly loaded Statim cassette.
Please refer to this operator's manual for instructions on the proper arrangement of
instruments in the cassette and the use of Stat-Dri plates (Statim 5000S). By carefully
following these directions on how to properly load the cassette chamber, rapid drying of the
load will be achieved.
4