The chamber door is open and
locked.
The cartridge door does not
open.
The cartridge door will not close.
The probe will not sit correctly in
the chamber.
The cycle will not start.
Liquid is leaking from the
trophon2 device.
The trophon2 device is failing
multiple cycles.
The trophon2 device takes too
long to warm up between cycles
• Power the trophon2 device OFF and back ON using the
power switch. Refer to Figure 4. The trophon2 device should
then automatically unlock the chamber door.
• If the above are not successful, power OFF the trophon2
device and follow SECTION F2: Manual Door Lock Override
• There is power to the trophon2 device.
• A HLD cycle is not in progress.
• The waste drawer is empty and fully inserted.
• There is nothing obstructing the cartridge door.
Note: The cartridge door is automatic, and should not be forced
open.
• Correct cartridge type has been inserted.
• Cartridge is correctly positioned.
• Cartridge lid has been removed.
• The probe is compatible for use in the trophon2 device – see
section A1.2: WarningsThe probe is loaded correctly.
• The IPP is correctly engaged or released – see Section C3.3
Positioning the Probe
• The probe is loaded correctly.
• The chamber door is closed.
• Confirm probe is clean & dry before pressing start.
• All onscreen instructions have been followed correctly.
WARNING: Any fluid leaking from the trophon2
device may contain hydrogen peroxide.
If liquid or mist is seen coming from the trophon2 device at any
time:
• Do not come into contact with the mist or liquid.
• Wear appropriate PPE.
• Ensure area is well ventilated.
• Allow the trophon2 device to complete the cycle.
• Turn off the trophon2 device and remove the power cord.
• Contact your customer service representative.
• Consult the trophon NanoNebulant SDS.
• Record any error codes and color of the trophon Chemical
Indicator and contact your customer service representative.
• Ensure that the probe is removed promptly at the end of each
cycle.
32
L03082 1.0 09/2021