Descargar Imprimir esta página

Integra MAYFIELD A2600R Manual De Instrucciones página 17

Adaptador para camilla espinal

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Cleaning of the MAYFIELD Spine Table Adaptor (A2600R, A2600M)
Do Not Steam Sterilize! The components may be damaged by heat.
Following these steps is recommended:
1. Remove Spine Table Adaptor from operating table parallel rails.
2. Remove Swivel Adaptor, and Link Arm Transitional Assembly from adaptor.
The MAYFIELD Base Handle must never be in the locked position
without a transitional member (or link arm) in its receptacle.
Failure to do so may result in damage to the handle receptacle.
3. The Spine Table Adaptor should be thoroughly cleaned after each use. Scrub each component
with a soft brush and use only a pH-neutral detergent. Clean thoroughly to remove any traces
of blood and/or debris and to prevent such blood or debris from interfering with function or
movement. Rinse thoroughly with clean water.
4. Dry all components with a soft dry towel.
5. After components are totally dry, re-assemble the components. Unit is ready for next use.
Optional Automatic Wash Cycle:
Any deviation from this guideline could result in damage to equipment as well as
improper cleaning results. Parameters given below must be followed exactly in order
to assure proper cleaning and avoid damage to the equipment.
Disassemble the device as indicated in section Cleaning of the MAYFIELD Spine Table Adaptor.
Rinse components under warm tap water prior to placing in washer.
Arrange device in a way to avoid contact between components.
Select instrument cycle and ensure proper programming.
WARNING!
CAUTION!
CAUTION!
17
EN – ENGLISH

Publicidad

loading

Este manual también es adecuado para:

Mayfield a2600m